fbpx

MR8740T

Max. 108 Analog Channels, Reduce Inspection Data Transfer Time to Zero

CONTACT US

Description

The MR8740T (MR8740-50) is rack-mountable data acquisition system that delivers high-speed, multi-channel measurements over maximum of 108 channels. Generate simulated signals and record at the same time using plug-and-play input units, making the waveform recorder ideal for evaluating multi-signal devices such as vehicle electronic control units (ECU).

Key Features
• Ideal for multipoint inspection of high performance boards such as ECU
• 108ch analog to 96ch analog + 48ch logic input
• Reduce time required to save to external media to max.1/100 compared with conventional method
• 20 MS/s simultaneous sampling on all channels
• Safe measurement with all analog inputs isolated
• Supports 4K monitor to display multi-channel waveforms without overlapping
• Measure 4 channels with 1 unit (4 ch analog Unit U8975, 4 ch DVM Unit U8991)
• Generate constant voltage, constant current, and simulated resistance (VIR Generator Unit U8794)

Model No. (Order Code)

MR8740-50Max. 108ch, 1GW memory, main unit only
LAN
CE

 

ECU Testing

ECUs are connected to a large number and wide variety of sensors. The MR8740T can be installed with signal generation unit to simulate these sensors. By measuring the simulation results with a measurement unit at the same time, you can perform all steps from signal generation to measurement with a single device. The U8794 VIR GENERATOR UNIT also offers resistance output to enable thermistor circuit testing.

Ideal for testing that requires simulated signals

When used as an ECU testing device, generate simulated signals from various sensors, which is indispensable for testing electronic parts and maintaining equipment. Record signals captured by various sensors to identify air flow, throttle, torque, speed, vibration, humidity, etc.

Basic specifications (Accuracy guaranteed for 1 year, Post-adjustment accuracy guaranteed for 1 year)

Number of input unitsMax. 27 slots
Number of channels[Using the U8975] Max. 108 ch analog, or 96 ch analog + 48 ch logic (when used in combination with U8975 + 8973)
[Using the 8966] Max. 54 ch analog, or 48 ch analog + 48 ch logic (when used in combination with 8966 + 8973)
*Logic unit 8973 is limited to slots 25 to 27, up to 3 units.
*Analog unit channels are isolated from each other and from chassis. Logic unit channels share a common GND with chassis.
Measurement ranges100 mV to 400 V f.s., 12 ranges, resolution : 1/2000 of range (when using 8966)
4 V to 200 V f.s., 6 ranges, resolution : 1/32000 of range (when using U8975)
100 mV to 1000 V f.s., 5 ranges, resolution : 1/1000 000 of range (when using MR8990)
1 V, 10 V, 100 V f.s., 3 ranges, resolution : 1/1000 000 of range (when using U8991)
Max. allowable input400 V DC (when using 8966; upper limit voltage that can be applied between input terminals without damage)
Max. rated voltage to earth300 V AC/DC (input and instrument are isolated; between input channels and chassis; upper limit voltage that can be applied between input channels without damage)
Frequency characteristicsDC to 5 MHz (-3 dB, when using 8966)
Max. sampling speed20 MS/s, all ch simultaneous, external sampling: 10 MS/s
Measurement functionsMemory (high-speed recording)
Memory capacityTotal of 1 G Word installed, 16 MW/ch (when using 8966), 8 MW/ch (when using U8975 or MR8990), 4 MW/ch (when using U8991)
Internal storageSSD 480 GB
Removable storageUSB memory stick ×8
Monitor outputVGA, HDMI, Display Port, Recommended resolution 1920 × 1080 dot or more
External interfaces[LAN] 1000 BASE-T, 100 BASE-TX, 10 BASE-TX (2 port) (DHCP and DNS support, FTP server/cliant, HTTP server)
[USB] USB 3.0 Series A receptacle × 4, USB 2.0 ×4
Power supply100 to 240 V AC, 50/60 Hz (400 VA max.)
Dimensions and mass426 mm (16.77 in)W × 177 mm (6.97 in)H × 505 mm (19.88 in)D, 14.0 kg (493.8 oz) (main unit only)
AccessoriesPower cord ×1,Quick Start Manual (booklet) ×1, Instruction Manual (detailed edition) (CD-R) ×1, application disk (CD-R) ×1, blank panel (blank slot only), rack installation hardware

Catalogs

HIGH ACCURACY / WIDEBAND CURRENT SENSOR Series CT686x, CT687x, CT684x

Catalog: MEMORY HiCORDER MR8740T

Technical Notes

Memory HiCorder MR8740T,Digital Voltmeter Unit U8991,VIR Generator Unit U8794

Declaration of Conformity

MEMORY HiCORDER MR8740-50

Drivers, Firmware

MATLABscript for MEMORY HiCORDERs (8860-50(51) 8861-50(51) MR8827 MR8847(A) MR8875 MR8740 MR8741 MR6000 MR8740T)

Wave Viewer (Wv)

LabVIEW driver for the MR8740T MEMORY HiCORDER

MEMORY HiCORDER & MEMORY HiLOGGER Application Disk

MR8740T(MR8740-50)

Options

10 mA class to 500 A (High speed)

 

CLAMP ON PROBE 3275

• DC to 2MHz (-3dB) • 500Arms maximum input • 0.01V/A output • φ 20 mm (0.79 in) core dia.

CLAMP ON PROBE 3274

• DC to 10MHz (-3dB) • 150Arms maximum input • 0.01V/A output • φ 20 mm (0.79 in) core dia.

CLAMP ON PROBE 3276

• DC to 100MHz (-3dB) • 30Arms maximum input • 0.1V/A output • φ 5 mm (0.20 in) core dia.

CLAMP ON PROBE 3273-50