Product Details
DT4281 DMM FMI
DMM Field Measurement Unit







Download Brochure
Download Brochure
Category:DMM

Description:
Speedy Performance of Professional Testing - Two Models Provide High Accuracy and Fast Response



•60,000 count, 5-digit display, high-resolution measurements

•±0.025% DC V basic accuracy, wide 20 Hz to 100 kHz AC V frequency characteristics

•Low-pass filter cuts high harmonics (when measuring inverter fundamental waveforms)

•Includes multiple measurement functions such as DC+ACV, temperature, capacitance, and frequency

•Includes terminal shutter mechanism for accident prevention (prevents erroneous test lead insertion)

•Measures large currents with optional clamp probe (only for DT4281, which has no 10 A terminal for accident prevention)

•USB communications function supports PC measurements (optional)

•Broad -15 (5°F) to 55°C (131°F) operating temperature range


Specifications:
Basic specifications (Accuracy guaranteed for 1 year)




DC Voltage 60.000 mV to 1000.0 V, 6 ranges, Basic accuracy: ±0.025 % rdg. ±2 dgt.

AC Voltage

45 to 65 Hz 60.000 mV to 1000.0 V, 6 ranges, Frequency characteristics: 20 Hz - 100 kHz

Basic accuracy: ±0.2 % rdg. ±25 dgt. (True RMS, crest factor 3)

DC + AC Voltage

45 to 65 Hz 6.0000 V to 1000.0 V, 4 ranges, Frequency characteristics: 20 Hz - 100 kHz

Basic accuracy: ±0.3 % rdg. ±30 dgt. (True RMS, crest factor 3)

Resistance 60.000 Ω to 600.0 MΩ, 8 ranges, (Conductance: 600.00 nS, DT4282 only)

Basic accuracy: ±0.03 % rdg. ±2 dgt.

DC Current 600.00 μA to 600.00 mA, 4 ranges, Basic accuracy: ±0.05 % rdg. ±5 dgt.

AC Current

45 to 65 Hz 600.00 μA to 600.00 mA, 4 ranges, Basic accuracy: ±0.6 % rdg. ±5 dgt. (True RMS, crest factor 3)

Frequency characteristics: 20 Hz - 20 kHz (at 600 μA to 600 mA range)

AC Current

(use with the Clamp on probe)

45 to 1k Hz 10.00 A to 1000 A, 7 ranges

Add the Clamp on probe accuracy to Basic accuracy: ±0.6 % rdg. ±2 dgt.

(True RMS, crest factor 3)

Peak DC V measurement: Signal width 4 msec or more (single), 1 msec or more (repeated) AC V, DC/AC A measurement:

Signal width 1 msec or more (single), 250 μsec or more (repeated)

C (capacitance) 1.000 nF to 100.0 mF, 9 ranges, Basic accuracy: ±1.0 % rdg. ±5 dgt.

Continuity Check Continuity threshold: 20/50/100/500 Ω, Response time: 10 ms or more

Diode test Open terminal voltage: 4.5 V or less, Testing current 1.2 mA or less, Threshold of forward voltage: 0.15 V to 3 V, seven stages

Frequency AC V, DC+AC V, AC A measurement, at pulse width 1 μs or more (50 % duty ratio)

99.999 Hz (0.5 Hz or more) to 500.00 kHz, 5 ranges, ±0.02 % rdg. ±3 dgt.

dB conversion Standard impedance setting (dBm), 4 Ω to 1200 Ω, 20 stages

Display dB conversion value of AC voltage (dBV)

Temperature K: -40.0 °C to 800.0 °C (-40.0 °F to 1472.0 °F) (use with optional probe DT4910)

Add accuracy of the Thermocouple probe to main unit accuracy: ±0.5 % rdg. ±3 °C

Other functions Filter function (Remove harmonic noise, use only at 600 VAC, 1000 VAC ranges), Display value hold, Auto hold, Max/Min value display, Sampling select, Relative display, Measurement memory (400 data), Auto-power save, USB communication (option), 4-20 mA % conversion

Display Main and Sub displays: 5-digits LCD, max. 60000 digits

Display refresh rates 5 times/s (Capacitance measurement: 0.05 to 2 times/s, depending on measured value, Temperature: 1 time/s )

Power supply LR6 (AA) alkaline batteries ×4, Continuous use: 100 hours

Dimensions and mass 93 mm (3.66 in)W × 197 mm (7.76 in)H× 53 mm (2.09 in)D, 650 g (22.9 oz) (with test leads holder and batteries)

Accessories TEST LEAD L9207-10 ×1, Instruction Manual ×1, LR6 Alkaline Battery ×4