Product Details
8423 Memory HiLogger
Up to 600 Channel Data Logger







Download Brochure
Download Brochure
Category:Memory Recorders/Loggers

Description:
Fast 10-ms Sampling. Up to 600 Channels Data Logging


Capture data with 15 to a maximum of 600 channels

Send data to the PC in real time

Isolated to sustain up to 600 V between modules and earth

USB 2.0, LAN 100BASE-TX, Store to 1GB PC Card

Simultaneous fast- and low-speed sampling allows for media storage space efficiency


Specifications:
Basic specifications (Accuracy guaranteed for 1 year)


No. of connectable units Maximum 8 units (total 120 channels), Bundle 8 Modules together to achieve a 120-channel System, Bundle 5 Systems together to enable a maximum of 600 channels of simultaneous recording

Measurement parameters

Model 8948 Voltage : 150mV to 100V, 1-5V f.s. 6 ranges, Max. resolution 5V

Temperature (thermocouples) : -200C to 2000C (depend on the sensor), 3 range (K, E, J, T, N, W (Wre5-26), R, S, B), Max. resolution 0.01C

Max. allowable input : DC 100V, Max. rated voltage between channels : 200 V DC, Max. rated voltage to earth : 600 V DC, AC

Measurement parameters

Model 8949 Voltage : 150mV to 60V, 1-5V f.s. 6 range, Max. resolution 5V

Temperature (thermocouples) : -200C to 2000C (depend on the sensor), 3 range (K, E, J, T, N, W (Wre5-26), R, S, B), Max. resolution 0.01C

Temperature (Resistance temperature sensor) : -200C to 800C, 3 range (Pt 100, JPt 100), Max. resolution 0.01C

Humidity : 100% rh f.s., 5.0 to 95.0% rh (use with the 9701), resolution 0.1% rh

Max. allowable input : DC 60V, Max. rated voltage between channels : 120 V DC, Max. rated voltage to earth : 600 V DC, AC

Measurement parameters

Model 8996 Totalized pulses : 0 to 1000M pulse, 1 range (No-voltage 'a' contact, open collector or voltage input), Max. resolution 1 pulse

Rotation count : 0 to 5000/n (r/s) f.s. 1 range (No-voltage 'a' contact, open collector or voltage input), Resolution 1/n (r/s) Note: n = pulses per rotation (1 to 1,000)

Digital input : ON/OFF digital signal

Max. allowable input : DC 50V, Max. rated voltage between channels : 33V AC, 70V DC, Max. rated voltage to earth : 600 V DC, AC, (Upper limit voltage that does not cause damage when applied between CH-1 to CH-5 each channel and chassis, CH-6 to CH-10 each channel and chassis, CH-11 to CH-15 each channel and chassis, and between each UNITs), (common ground for CH-1 to CH-5, common ground for CH-6 to CH-10, common ground for CH-11 to CH-15)

Recording intervals 10ms to 1hr, 19 range (5s to 1hr when combined with humidity measurement), Dual sampling : Recording intervals can be specified for every input module (high-speed and low-speed)

Function Measurement data are saved to the CF Card in real time, Trigger function, Digital filter (Input unit), Alarm output (use with the ALARM UNIT 8997), Data acquisition is controlled by the PC data acquisition program, FTP server function, HTTP server function

Interface LAN : supports 100Base-TX, USB : Ver 2.0, mini-B receptacle, CF card slot

Power supply Using the AC ADAPTER 9418-15 /20 VA (when connected with 8 units), 12V Battery (voltage may range from -20% to +30%, Please contact HIOKI for connection cord).

Dimensions & Mass 67 mm (2.64 in) W 133 mm (5.24 in) H 125 mm (4.92 in) D, 600 g (21.2 oz) (main unit 8423 only)

Accessories Operating Manual 1, Quick Start Manual 1, AC ADAPTER 9418-15 1, USB cable 1, Connection Plate 1, CD-R (data collection software "LOGGER UTILITY") 1, Connector cover 1, Ferrite clamp 1