3-Phase 4-Wire Power Quality Analyzer Compliant to IEC61000-4-30 Class A Standards with Dynamic 1300V Range, Harmonic and Inter-harmonics Analysis Capabilities

SKU: PW3198 Category:

Model No. (Order Code)

PW3198 main unit, clamp sensor is sold separately
PW3198-90 kit includes software the 9624-50

Note: Voltage can be measured with the main unit alone. An optional current sensor is necessary to measure current or power parameters. Use the PQA-HiView Pro 9624-50 (version 2.00 or later) with a PC to analyze the data collected to the SD card.

LAN
USB 2.0
CE
TrueRMS
RS232C
SD

 

Description

Hioki portable power meters and analyzers are best in class power measuring instruments for measuring single to three-phase lines with a high degree of precision and accuracy. The PW3198 is a power quality analyzer for monitoring and recording power supply anomalies, allowing their causes to be quickly investigated, and also for assessing power supply problems such as voltage drops, flicker, harmonics, and other electrical issues.

Key Features
• Verify power problems in accordance with the IEC61000-4-30 Class A standard
• High accuracy and continuous gapless recording
(V: ±0.1% of nominal voltage, A and W: ±0.2% rdg. ±0.1% f.s.)
• CAT IV 600V - safe enough for incoming power lines
• Broadband voltage range lets you measure even high-order harmonic components of up to 80 kHz
• Wide dynamic range from low voltages up to 1300V (3P4W line-to-line voltage)
• Maximum 6000V transient overvoltage up to 700kHz
• LAN, USB and SD card interfaces
• Optional GPS BOX for synchronizing multiple devices

Product Video

 

Basic specifications (Accuracy guaranteed for 1 year, Post-adjustment accuracy guaranteed for 1 year)

Measurement line type Single-phase 2-wire, Single-phase 3-wire, Three-phase 3-wire or Three-phase 4-wire plus one extra input channel (must be synchronized to reference channel during AC/DC measurement)
Voltage ranges Voltage measurement: 600.00 V rms
Transient measurement 6.0000 kV peak
Current ranges 500.00 mA to 5.0000 kA AC (depends on current sensor in use)
Power ranges 300.00 W to 3.0000 MW (determined automatically based on voltage and current range in use)
Basic accuracy Voltage: ±0.1% of nominal voltage
Current: ±0.2 % rdg. ±0.1 % f.s. + current sensor accuracy
Active power: ±0.2 % rdg. ±0.1 % f.s. + current sensor accuracy
Measurement items 1. Transient over voltage : 2 MHz sampling
2. Frequency cycle : Calculated as one cycle, 40 to 70 Hz
3. Voltage (1/2) RMS: one cycle calculation refreshed every half cycle
Current (1/2) RMS: half-cycle calculation
4. Voltage swell, Voltage dips, Voltage interruption
5. Inrush current
6. Voltage waveform comparison
7. Instantaneous flicker value: As per IEC61000-4-15
8. Frequency: Calculated as 10 or 12 cycles,40 to 70 Hz
9. 10-sec frequency: Calculated as the whole-cycle time during the specified 10 s period, 40 to 70 Hz
10. Voltage waveform peak, Current waveform peak
11. Voltage, Current, Active power, Apparent power, Reactive power, Active energy, Reactive energy, Power factor, Displacement power factor, Voltage unbalance factor, Current unbalance factor (negative-phase, zero-phase)
12. High-order harmonic component (voltage/ current): 2 kHz to 80 kHz
13. Harmonic/ Harmonic phase angle (voltage/ current), Harmonic power: 0th to 50 th orders
14. Harmonic voltage-current phase angle: 1th to 50 th orders
15. Total harmonic distortion factor (voltage/ current)
16. Inter harmonic (voltage/ current): 0.5 th to 49.5 th order
17. K Factor (multiplication factor)
18. IEC Flicker, Δ V10 Flicker
Record 55 weeks (with repeated recording set to [1 Week], 55 iterations)
35 days (with repeated recording set to [OFF])
Interfaces SD/SDHC card, RS-232C, LAN (HTTP server funtion), USB2.0
Display 6.5-inch TFT color LCD (640 × 480 dots)
Power supply AC Adapter Z1002 (12 V DC, Rated power supply 100 V AC to 240 V AC, 50/60 Hz)
Battery Pack Z1003 (Ni-MH 7.2 V DC 4500 mAh)
Dimensions and mass 300 mm (11.81 in) W × 211 mm (8.31 in) H × 68 mm (2.68 in) D (excluding protrusions), 2.6 kg (91.7 oz) (including battery pack)
Accessories Instruction manual ×1, Measurement guide ×1, Voltage cord L1000 ×1 set (Red/Yellow/Blue/Gray, each 1, Black ×4, Alligator clip ×8), Spiral tube ×20, Input Cable Labels ×1, AC adapter Z1002 ×1, Strap ×1, USB cable (1 m 3.28 ft length) ×1, Battery pack Z1003 ×1, SD memory card 2GB Z4001 ×1

Catalogs

Catalog: PQA-HiVIEW PRO 9624-50

POWER QUALITY ANALYZER PW3198

Declaration of Conformity

POWER QUALITY ANALYZER PW3198 [111KB]

Others

Measuring Instruments for the Automotive Market

Options

500 A to 5000 A

*For commercial power lines, 50/60 Hz

 

AC FLEXIBLE CURRENT SENSOR CT9667

• 10 Hz to 20 kHz (±3dB) • 5000 A/ 500 A AC rated input • 500 mV/f.s. output • φ 100 mm (3.94 in) core dia. to φ 254 mm (10.00 in), 3 loop diameters

Adapter

 

CLAMP ON ADAPTER 9290-10

• Primary 1000A, secondary 100A (1/10 ratio) output • Superior phase angle characteristics for power

Cases

 

CARRYING CASE C1001

Soft type, Includes compartment for options

CARRYING CASE C1002

For the PW3198, hard trunk type, for storing options

Current input

 

CLAMP ON SENSOR 9694

• 40Hz to 5kHz • Phase: 45Hz to 5kHz • 5 A AC rated input • 10 mV AC / A output • φ 15 mm (0.59 in) core dia.

CLAMP ON SENSOR 9695-03

• 40 Hz to 5 kHz • Phase: 45 Hz to 5 kHz • 100A AC rated input • 1mV AC/A output • φ 15 mm (0.59 in) core dia.

CLAMP ON SENSOR 9661

• 40Hz to 5kHz • Phase: 45Hz to 5kHz • 500A AC rated input • 1mV AC/A output • φ 46 mm (1.81 in) core dia.

CLAMP ON SENSOR 9660

• 40Hz to 5kHz • Phase: 45Hz to 5kHz • 100 A AC rated input • 1 mV AC / A output • φ 15 mm (0.59 in) core dia.

Current sensors (AC leak)

 

CLAMP ON LEAK SENSOR 9675

• 40 Hz to 5 kHz • Rated 10 A AC • 100 mV AC/A output • φ 30 mm (1.18 in) core dia.

CLAMP ON LEAK SENSOR 9657-10

• 40 Hz to 5 kHz • Rated 10 A AC • 100 mV AC/A output • φ 40 mm (1.57 in) core dia.

Current sensors (AC load)

 

CLAMP ON SENSOR 9669

• 40Hz to 5kHz • Phase: 45Hz to 5kHz • 1000A AC rated input • 0.5mV AC/A output • φ55 mm (2.17 in) core dia.

CLAMP ON SENSOR 9695-02

• 40 Hz to 5 kHz • Phase: 45 Hz to 5 kHz • 50A AC rated input • 10mV AC /A output • φ 15 mm (0.59 in) core dia.

CONNECTION CABLE 9219

Connect with the 9695-02/ -03, Output BNC terminal, 3m (9.84ft) length

Input cable (B)

*Voltage is limited to the specifications of the input modules in use.

 

GRABBER CLIP 9243

Attaches to the tip of the banana plug cable, Red/ Black: 1 each, 196 mm (7.72 in) length, CAT III 1000 V

Options

 

MAGNETIC ADAPTER 9804-02

Attaches to the tip of cord, black ×1, φ11 mm (0.43 in)

Other options

 

PQA-HiVIEW PRO 9624-50

• For PW3198, 3197, 3196 • Easy operation, analyze data on the PC • Convenient report creation function

GPS BOX PW9005

To synchronize the PW3198 clock to UTC

PC Software

 

LAN CABLE 9642

Straight Ethernet cable, supplied with straight to cross conversion adapter, 5 m (16.41 ft) length

Power supply

*Not necessary when using Current Unit 8971.

 

AC ADAPTER Z1002

For main unit, 100 to 240 V AC

BATTERY PACK Z1003

NiMH, Charges while installed in the main unit

Storage media

*The CF card includes a PC card adapter. PC Card Precaution Use only PC Cards sold by HIOKI. Compatibility and performance are not guaranteed for PC cards made by other manufacturers. You may be unable to read from or save data to such cards.

 

SD MEMORY CARD 2GB Z4001

2 GB capacity

Voltage Input

 

VOLTAGE CORD L1000

1000 V specifications, Red/ Yellow/ Blue/ Gray each 1, Black 4, Alligator clip ×8, 3m (9.84ft) length

WIRING ADAPTER PW9000

For three-phase 3-wire

WIRING ADAPTER PW9001

For three-phase 4-wire

Voltage measurement

*PW9020 ×4 pieces are bundled, additional single sensors also available

 

MAGNETIC ADAPTER 9804-01

Attaches to the tip of cord, red ×1, φ11 mm (0.43 in)

Measure Transient Overvoltage Waveforms 

Measure Periodical Instantaneous Voltage Drops 

Measure Voltage Waveforms when Switching to UPS 

Inspect the Stress on Factory UPS Caused by Voltage Dips 

Checking the Amount of Electricity Sold in Photovoltaic (PV) Systems 

Measure the High-Order Harmonic Voltage and Current of Power Lines 

Power Quality Survey of Solar Power Generation Systems 

Measure the Noise Generated by Switching Power Factor Improvement Capacitors 

Measure the Transient Overvoltage in Glowing Fluorescent Lights 

Measure Lightning-induced Voltage Dips in a Low-Voltage Circuit 

Measure Voltage Drop Caused by Wiring Impedance 