fbpx

MEMORY HiLOGGER LR8400 Series

Home/Multichannel Data Loggers/MEMORY HiLOGGER LR8400 Series

MEMORY HiLOGGER LR8400 Series

Portable Handheld 30- to 60- channel Data Loggers with 10ms Simultaneous Sampling

SKU: LR8400 Series Category:

Model No. (Order Code)

LR8400-20 built-in the Voltage/temp unit LR8500 ×2, 30 ch
LR8401-20 built-in the Universal unit LR8501 ×2, 30 ch
LR8402-20 built-in the Universal unit ×1, Voltage/temp unit ×1, 30 ch

Note: Caution: Built-in units cannot be removed or changed. The Battery pack Z1000 is sold separately Model LR8400 : Built-in units are equivalent to the Votage/temp unit LR8500 ×2 Model LR8401 : Built-in units are equivalent to the Universal unit LR8501 ×2 Model LR8402 : Built-in units are equivalent to the Universal unit LR8501 (15 ch) ×1, and the Votage/temp unit LR8500 (15 ch) ×1

LAN
USB 2.0
CE

 

Description

Hioki Memory HiLoggers are high-speed data loggers for recording multiple chanels of voltage, temperature, resistance, humidity and pulse signals, providing complete isolation between channels and strong noise resistance. The LR8400 Series logs data at 10ms speeds across 30 channels simultaneously, is expandible up to 60 channels, and features a bright color LCD for crystal clear waveform displays.

Key Features
• Compact size despite 30-channel standard capabilities
• Expand up to 30 additional channels
• Protected against unexpected power outages
• Write data to USB memory stick or CF card in real-time
• Built in with USB 2.0 or 100 BASE-TX LAN interfaces
• 5.7" TFT color display

Product Video

 

A multichannel data logger that’s ideal for battery, automotive, and environmental measurement

For applications such as
● Development of fuel cell materials, energy field
● Helps also in collecting automotive data
● Maintenance and inspection of equipment
● Testing of electrical products

 

 

Portable data logger with 30 standard channels expandible to 60 channels

The need for more measurement channels can be met even after purchasing the instrument.

The instrument comes with 30 channels as standard, but another two 15 channel input units can be added to expand the total number of channels to 60.
(All 30 analog channels are isolated.)

The instrument also features 8 channels of pulse and digital input.
Each channel can configured for pulse or digital input as desired.

Note: The units provided with the unit as standard cannot be removed.

 

 

A data logger with a sampling speed of up to 10 ms

The development of hybrid and electric automobiles requires instruments that can measure abrupt load changes.

The instrument can reliably capture even abrupt changes in load since it can measure channels 1 to 15 in 10 ms.
Channels 16 to 30 provide 20-ms sampling.

Note: Measurements on channels 31 to 60 provide 50-ms sampling.

 

 

A high-withstand-voltage data logger with a maximum rated channel-to-channel and channel-to-ground voltage of 300 V

The HiLOGGER measures not only fuel cells, but also batteries for UPS (uninterruptible power supplies) devices used in buildings as well as batteries consisting of cells and packaging connected in stacks that require multi-point measurements.

In such measurements, high voltage for the whole stack is applied between channel-to-channel and channel-to-ground.

The universal unit delivers safe measurement thanks to a high-withstand-voltage design featuring a maximum rated channel-to-channel voltage of 300 V DC.
(Maximum rated voltage between each channels of VOLTAGE/TEMP UNIT: 250 V DC)

Note: Isolation between channels is possible through the use of semi-conductor relays.
Voltage exceeding the product specifications, such as that originating from lightning surges or other sources, should never be applied between each channel; otherwise the relays will short and the recorder will be damaged.

 

 

Included PC application with measurement and analysis functionality

The supplied Logger Utility software enables processing of measurement data on a PC.

Use the supplied Logger Utility software to record data on a PC in real-time.
View past data during recording.

Connect the PC to the HiLOGGER using USB or via LAN.

 

 

Basic specifications (Accuracy guaranteed for 1 year, Post-adjustment accuracy guaranteed for 1 year)

Analog input Universal unit LR8501 Actual functionality will depend on combination of units installed Note: Isolated from each channel to chassis [No. of channels] 15 analog channels; isolated scanning method input (4 terminals: push-button type)
[Voltage] ±10 mV to ±100 V, 1-5 V f.s. Max. resolution: 500 nV, (Isolated between channels and from each channel to chassis)
[Temperature: Thermocouples] -200 ˚C to 2000 ˚C (depends on sensor), Thermocouples (K, J, E, T, N, R, S, B, W,) Max. resolution 0.01 ˚C, (Isolated between channels and from each channel to chassis)
[Temperature: Pt 100, JPt 100 sensor] -200 ˚C to 800 ˚C, Max. resolution 0.01 ˚C, (Not isolated between channels)
[Resistance] 0 Ω to 200 Ω f.s. Max. resolution 0.5 mΩ, (Not isolated between channels)
[Humidity] 5.0 to 95.0 % rh (use with optional sensor), resolution 0.1 % rh, (Not isolated between channels nor from each channel to chassis)
[Max. rated voltage between isolated input channels] 300 V DC
[Max. allowable input] ±100 V DC
[Max. rated voltage from isolated terminals to ground] 300 V AC, DC
Analog input Votage/temp unit LR8500 Actual functionality will depend on combination of units installed Note: Isolated from each channel to chassis [No. of channels] 15 analog channels; isolated scanning method input (2 terminals: M3 screw type)
[Voltage] ±10 mV to ±100 V, 1-5 V f.s. Max. resolution: 500 nV, (Isolated between channels and from each channel to chassies)
[Temperature: Thermocouples] -200 ˚C to 2000 ˚C (depends on sensor), Thermocouples (K, J, E, T, N, R, S, B, W), Max. resolution 0.01 ˚C, (Isolated between channels and from each channel to chassies)
[Pt 100, JPt 100 sensor] [Resistance] Not available
[Humidity] 5.0 to 95.0 % rh (use with optional sensor), resolution 0.1 % rh, (Not isolated between channels nor from each channel to chassies)
[Max. rated voltage between isolated input channels] 250 V DC
[Max. allowable input] ±100 V DC
[Max. rated voltage from isolated terminals to ground] 300 V AC, DC
Caution: Max. voltage from terminals to ground without damage
Pulse, Digital input [No. of channels] 8 channels, pulse / digital selectable for each channel, M3 screw terminal, not isolated, common ground
[Pulse totalization] 0 to 1000 M pulse, 1 range (No-voltage 'a' contact; normally open, open collector or voltage input), Max. resolution 1 pulse
[Rotation count] 0 to 5000 /n (r/s) f.s. 1 range (same as Pulse totalization input signal condition), resolution 1/n (r/s) Note: "n" is the number of sensor output pulses per revolution, 1 to 1000
[Digital input] Record logical "1" or "0" at each sampling
[Max. rated voltage between input channels] Not isolated
[Max. allowable input] 0 to 50 V
Recording intervals 10 ms to 50 ms, 100 ms to 1 hour, 19 selections (All input channels are scanned within each recording interval.)
Note: limited by using channels at 10 ms to 50 ms interval
Digital filter Select from OFF/ 50 Hz/ 60 Hz (the cut-off frequency is automatically set)
Data storage Internal memory: 8 M-words, Data storage media: CF card or USB memory (Only data recorded to a genuine HIOKI CF card is guaranteed)
LAN interfaces 100BASE-TX, Functions: Data acquisition using bundled software or PC commands, FTP server, FTP client, HTTP server function, or E-mail system
Note: LAN communication support planned from software Ver. 1.20
USB interfaces USB 2.0 High-speed capable, series mini-B receptacle
Functions: Data acquisition using bundled software or PC commands, Transfer data from the CF card to a PC via USB drive mode
Note: Data transfer not possible from USB memory sticks
Display device 5.7 inch TFT color liquid crystal display (640 × 480 pixel)
Other functions Save waveform data in real time to the CF card or USB memory stick,
Numerical value calculations, Waveform calculations, and others
AC Power supply Using the AC adapter 9418-15 (100 to 240 V AC, 50/60 Hz), 7 VA
DC Power supply Using the Battery pack Z1000 (optional accessory), Continuous use 5 hr,
External power: 10 to 28 V DC (Please contact your HIOKI distributor for connection cord)
Dimensions and Mass 272 mm (10.71 in) W × 182.4 mm (7.18 in) H × 66.5 mm (2.62 in) D,
1.8 kg (63.5 oz), (LR8400-20 main unit, excluding the Battery Pack 370 g/ 13.1 oz)
Accessories Instruction manual ×1, Measurement guide ×1, CD-R (data collection software "Logger Utility") ×1, USB cable ×1, AC adapter 9418-15 ×1

Combination status with LabVIEW or MATLAB script

Compatible models LabVIEW MATLAB script
LR8400-20 OK N/A
LR8401-20 OK N/A
LR8402-20 OK N/A

Catalogs

MEMORY HiLOGGER LR8400, LR8401, LR8402

Declaration of Conformity

MEMORY HiLOGGER LR8400-20,LR8401-20,LR8402-20 [100KB]

Others

Measuring Instruments for the Automotive Market

Options

Carrying cases and stands

 

CARRYING CASE C1000

For the LR8400 series, for storing options

FIXED STAND Z5000

For wall hanging and slanted bench mounting

For the PC

*LOGGER UTILITY is a bundled software

 

LOGGER UTILITY SF1000

• For the HIOKI Memory HiLoggers • The control of the measurement of loggers, real-time data collection • Wavefrorm display, data conversion, calculation, search, print

Input cable (B)

*Voltage is limited to the specifications of the input modules in use.

 

GRABBER CLIP 9243

Attaches to the tip of the banana plug cable, Red/ Black: 1 each, 196 mm (7.72 in) length, CAT III 1000 V

Input cable (D)

* Voltage to ground is within this product’s specifications, separate power source is also required.

 

AC ADAPTER Z1008

100 to 240V AC

DIFFERENTIAL PROBE P9000

• Waveform only for exclusive use with Memory HiCorders (P9000-01) or waveform/RMS value switchable (P9000-02) • For use with input of up to 1 kV AC/DC • Requires separate power supply

Input cable (E)

*Voltage to ground is within this product's specifications. Separate power source is also required.

 

AC ADAPTER 9418-15

100 to 240V AC

Input modules

*Input cords are not included. Please purchase them separately. *When using 9709 with Current Unit 8971, a total of 7 current probes can be used.

 

UNIVERSAL UNIT LR8501

4 terminals push-button type, 15 channels, Voltage, Temperature with thermocouple, Platinum Resistance temperature sensor, Humidity, or Resistance measurement, for the LR8400 series

VOLTAGE/TEMP UNIT LR8500

2 terminals M-3 mm screw type, 15 channels, Voltage, Temperature with thermocouple, or Humidity measurement, for the LR8400 series

Input options

 

HUMIDITY SENSOR Z2000

3 m (9.84 ft) length

Output cables

* Please contact your local HIOKI distributor for connectors that support Model MR8791.

 

CONVERSION CABLE L1011-10

2.4 m (7.87 ft) length, covert BNC to wire

CONVERSION CABLE L1011

30 cm (0.98 ft) length, covert BNC to wire

PC Software

 

LAN CABLE 9642

Straight Ethernet cable, supplied with straight to cross conversion adapter, 5 m (16.41 ft) length

Power supply

*Not necessary when using Current Unit 8971.

 

BATTERY PACK Z1000

NiMH, Charges while installed in the main unit

Storage media

*The CF card includes a PC card adapter. PC Card Precaution Use only PC Cards sold by HIOKI. Compatibility and performance are not guaranteed for PC cards made by other manufacturers. You may be unable to read from or save data to such cards.

 

PC CARD 512M 9728

• 512 MB capacity • PC card adapter bundled

PC CARD 9830

• 2 GB capacity • PC card adapter bundled

PC CARD 1G 9729

• 1 GB capacity • PC card adapter bundled

Test the Temperature Distribution of a Radiator 

Log 4 to 20 mA Current for Instrumentation, Distributors, etc. 

Logging Measurements using a Wireless Local Area Network 

Test Lumen Maintenance Factor According to the IES LM-80 Standard 

Measure the Temperature of Live Lighting Fixture Parts 

Test the Flexibility of Substrates and Signal Cables 

Measure the Voltage, Current, Temperature, and Humidity of a Fuel Cell Stack 

Drivers, Firmware, Software