INSULATION / WITHSTANDING HiTESTER 3159-02

Home/Electrical Safety Testers Hipot Testers/INSULATION / WITHSTANDING HiTESTER 3159-02

INSULATION / WITHSTANDING HiTESTER 3159-02

0 to 0.2 kV/5 kV AC Insulation and Hipot Tester

Model No. (Order Code)

3159-02 220 V AC power supply
RS232C
Not CE

 

Description

The Hioki electrical safety testers are designed for insulation resistance and voltage withstand testing of electrical devices and components according to various safety standards. The 3159 and 3159-02 tests insulation resistance at 500V or 1000V and withstand voltage of up to 500VA in sequence.

 

 

Key Features

• Continuous testing of insulation (500/1000 V) and withstand voltage (500 VA transformer capacity)
• Insulation to withstand series test or witstand to insulation series test at auto mode, or individual test at manual mode
• Save up to 10 test settings each for the withstanding and insulation testing modes
• External I/O, RS-232C interface, Status output (relay contacts)

 

 

Basic specifications (Accuracy guaranteed for 1 year, Post-adjustment accuracy guaranteed for 1 year)

[Withstanding test section]  
Testing voltage 0 to 2.5 kV / 0 to 5.0 kV AC, 2 range configuration 500 VA (30 minutes rated)
Voltage setting Manual setting
Waveform/Frequency Same as the power supply waveform, synchronized with the power supply
Current measurement 0.01 mA to 120 mA, True RMS rectified (digital display)
Voltage meter Accuracy: ±1.5 % f.s. (digital), ±5 % f.s. (analog, f.s.=5 kV)
Current measurement 0.01 mA to 120 mA, (Average value rectified, effective value digital display)
Measurement range 2 mA/8 mA (0.01 mA resolution), 32 mA (0.1 mA resolution), 120 mA (1 mA resolution)
Voltage meter Digital, Accuracy: ±1.5 % rdg. (f.s.=5.00 kV)
Judgment function Window comparator method (Digital setting)
[Insulation test section]  
Testing voltage 500 V DC, 1000 V DC, Unloaded voltage: 1 to 1.2 times rated voltage
Rated testing current 1 to 1.2 mA, Shorted current: 4 to 5 mA (at 500 V), 2 to 3 mA (at 1000 V)
Measurement range, Accuracy 0.5 MΩ to 999 MΩ (at 500 V), and 1 MΩ to 999 MΩ (at 1000 V): ±4 % rdg.,
1000 MΩ to 2000 MΩ: ±8 % rdg.
Judgment function Window comparator method (Digital setting)
[Timer section]  
Setting range 0.5 to 999 s
[General section]  
Monitor function Output voltage, detected current, insulation resistance, Refresh rate: 2 times/s
Power supply 220 V AC, (50/60 Hz), 800 VA max.
Dimensions and mass 320 mm (12.60 in)W × 155 mm (6.10 in)H × 330 mm (12.99 in)D,
21.5 kg (758.4 oz)
Accessories H.V. Test lead 9615 (high voltage side and return, 1 each) ×1, Power cord ×1, Instruction manual ×1, Spare fuse ×1

Catalogs

INSULATION / WITHSTANDING Test Instrument series 3153/3159/ST5520

User Guides

Introducing the safety standards, testing methods, and appropriate test instruments for electrical medical equipment

Options

Input/Output cords

*The 9615 is bundled

 

H.V. TEST LEAD 9615

Red, Black each 1, 1.5 m (4.92 ft) length

REMOTE CONTROL BOX (SINGLE) 9613

For Start/Stop control, 1.5m (4.92 ft) cord length

REMOTE CONTROL BOX (DUAL) 9614

For Start/Stop control, 1.5m (4.92 ft) cord length

Other

 

SAFETY TEST DATA MANAGEMENT SOFTWARE 9267

• PC software for controlling electrical safety measuring instruments

PC communication

 

RS-232C CABLE 9637

For the PC, 9 pin - 9 pin, cross, 1.8 m (5.91 ft) length

RS-232C CABLE 9638

For the PC, 9 pins - 25 pins, cross, 1.8m (5.91 ft) length

There are no drivers, firmware or software available for this product.