Compact & Lightweight Heat Flow Logger for Energy Conservation Analysis

SKU: LR8432 Category:

Model No. (Order Code)

LR8432-20 10 ch, english model

Note: The LR8432-20 is not bundled with the Battery Pack 9780. Thermocouples are not provided by HIOKI, and must be purchased from a separate vendor. Note: Use only HIOKI CF cards, which are manufactured to strict industrial standards, for long-term storage of important data. Correct operation of non-HIOKI CF cards or USB memory sticks is not guaranteed.

USB 2.0
CE

 

Description

Hioki Memory HiLoggers are high-speed data loggers for recording multiple chanels of voltage, temperature, pulse or rotation signals, providing complete isolation between channels and strong noise resistance. The Hioki LR8432 is a handheld 10-channel data logger especially designed for heat flow measurement. The Heat Flow Logger is ideal for evaluating insulation performance and analyzing the causes of temperature change.

Key Features
• Use a heat flow sensor to measure the movement and volume of heat energy
• Easily analyze temperature and heat flow
• Record measurement data on a USB flash drive for easy transfer to a computer
• Record to reliable Compact Flash cards during long-term measurement applications for increased peace of mind
• Replace storage media during real-time recording
• Improved thermocouple measurement accuracy and reference junction compensation accuracy
• Ten isolated analog input channels
• 10 ms sampling and recording across all channels
• Noise-resistant measurement circuitry for improved readings
• Ultra-compact for convenient portability
• Widescreen, bright LCD gives excellent viewability

Principal Applications
* Automotive market (evaluating the flow of heat into the vehicle interior from the engine compartment, assessing heat damage to vehicle parts, identifying sources of heat, etc.)
* Residential construction market (evaluating the insulation performance of building materials, evaluating the performance of radiant heating systems, etc.)

 

 

Hioki LR8432 Heat Flow Logger Explainer Video

Hioki Memory HiLoggers are high-speed data loggers for recording multiple chanels of voltage, temperature, pulse or rotation signals, providing complete isolation between channels and strong noise resistance. The Hioki LR8432 is a handheld 10-channel data logger especially designed for heat flow measurement. The Heat Flow Logger is ideal for evaluating insulation performance and analyzing the causes of temperature change.

 

 

A heat flow logger that’s ideal for addressing the root cause of changes in temperature

The LR8432-20 is a heat flow logger that visualizes quantities of heat.
By using a heat flow sensor to visualize movements of thermal energy (heat flows) and quantities of thermal energy, you can obtain data that serves as a leading indicator of changes in temperature.

 

 

Heat flow loggers as a useful tool for formulating thermal measures in R&D and performance evaluation

Visualizing heat flows (inflows and outflows) and quantities of heat with a heat flow logger is useful
when you need to implement more precise air-conditioning control or formulate thermal measures
as part of the product development process.

●Isolation of heat generated and dissipated in automotive parts
●Evaluate the thermal performance of building materials
●Measure the energy efficiency of consumer electronics
●Study the impact of body heat
●Diagnose the deterioration of insulation material in plant piping
●Index temperature fluctuation in agriculture and civil engineering

 

 

Display heat flow and temperature gauges simultaneously

Heat flow and temperature gauges can be displayed simultaneously for data
measured with the Heat Flow Logger LR8432-20.
Display the gauges for data you want to compare at the same time
in order to see changes in temperature and heat flow at a single glance.

 

 

Waterproof heat flow sensor that can measure curved surfaces

This heat flow sensor can measure curved surfaces such as those found on water pipes.
Choose from three available sizes to measure objects that were impossible to isolate for measurement before,
for example the surfaces of small parts and narrow gaps.
Sensor’s Operating temperature range:-40℃〜150℃ (-40°F to 302°F)

 

 

Basic specifications (Accuracy guaranteed for 1 year, Post-adjustment accuracy guaranteed for 1 year)

Specialized functions for heat flow measurement Easy scaling settings:directly enter the sensitivity of the heat flow sensor
Calculations:waveform processing function for the analysis of temperature and heat flow (Simple average, moving average, integration, heat transmission coefficient), Integration with numerical calculations
Analog inputs [No. of channels] 10 isolated analog channels using scanning input method (M3 mm dia. screw terminal block)
[Voltage measurement range] ±10 mV to ±60 V, 1-5V, Max. resolution 500 nV
[Temperature : thermocouples] −200 ˚C to 1800 ˚C (depending on sensor), thermocouples (K, J, E, T, N, R, S, B), Max. resolution 0.1 ˚C
[Humidity] not available
[Max. allowable input] 60 V DC
[Max. rated voltage between input channels] [Max. rated voltage to earth]
30 AC Vrms, 60 V DC (max. voltage between input channel terminals, and from terminals to chassis ground without damage)
Pulse inputs [No. of channels] 4 pulse input channels (requires CONNECTION CABLE 9641,
all pulse inputs share common ground with the main unit)
[Totalized pulses] 0 to 1000M (count) (No-voltage ‘a’ contact, open collector or voltage input), Max. resolution 1 pulse
[Rotation count] 0 to 5000/n (r/s), Resolution 1/n (r/s) * n = pulses per rotation (1 to 1,000)
[Max. allowable input] 0 to 10 V DC
[Max. rated voltage between input channels] [Max. rated voltage to earth] Non-isolated
Recording intervals 10 ms to 1 hour, 19 selections (All input channels are scanned at high speed during every recording interval)
Selectable filters 50 Hz, 60 Hz, or OFF (digital filtering of high frequencies on analog channels)
Memory capacity Internal storage: 3.5 M-words, External storage: CF card or USB memory stick (only the HIOKI CF cards are guaranteed for correct operation)
External interface USB 2.0 mini-B receptacle ×1; Functions: Control from a PC, Transfers files from the installed CF card to a PC (cannot transfer files from the connected USB memory stick to a PC via USB communication), Data copy between CF card and USB memory stick
Display 4.3-inch WQVGA-TFT color LCD (480 × 272 dots)
Functions Save data to the CF Card or USB memory stick in real time, Numerical Calculations, etc.
Power supply AC Adapter Z1005: 100 to 240 VAC (50/60 Hz), 30 VA Max. (including AC adapter), 10 VA Max. (main unit only)
Battery Pack 9780: Continuous use 2.5 hours (@25°C/77°F), 3 VA Max.
External power source: 10 to 16 V, 10 VA Max. (please contact HIOKI distributor for cable; less than 3 m/9.84 ft cable length)
Dimensions and mass 176 mm (6.93 in) W × 101 mm (3.98 in) H × 41 mm (1.61 in) D, 550 g (19.4 oz) (Battery Pack 9780 not installed)
Accessories Measurement Guide ×1, CD-R (Instruction manual PDF, Logger Utility Instruction Manual PDF, Data acquisition application program Logger Utility) ×1, USB cable ×1, AC Adapter Z1005 ×1

LabVIEW and MATLAB support

Compatible models LabVIEW MATLAB script
LR8432 N/A N/A

Catalogs

HEAT FLOW LOGGER LR8432

Technical Notes

Heat Flow Logger LR8432-20 and Heat Flow Sensor Z2012/Z2013/Z2014/Z2015/Z2016/Z2017

Declaration of Conformity

HEAT FLOW LOGGER LR8432-20 [101KB]

Others

A new proposal for customers who need to optimize thermal designs

Options

For the PC

*LOGGER UTILITY is a bundled software

 

LOGGER UTILITY SF1000

• For the HIOKI Memory HiLoggers • The control of the measurement of loggers, real-time data collection • Wavefrorm display, data conversion, calculation, search, print

Heat flow measurement

For measuring small electronic components or curved surfaces; Waterproof IP06, IP07

 

THERMALLY CONDUCTIVE TAPE Z5008

HEAT FLOW SENSOR Z2012 to Z2017

・Flexible sensor that can measure curved surfaces ・High sensitivity and high cost performance

Input cables

 

CONNECTION CABLE 9641

For pulse input, 1.5 m (4.92 ft) length

Other options

 

SOFT CASE 9812

Includes space for small items, neoprene rubber

CARRYING CASE 9782

Includes compartment for options, resin-coated

PROTECTION SHEET 9809

For LCD protection, pairs of sheets

Power supply

*Not necessary when using Current Unit 8971.

 

AC ADAPTER Z1005

100 to 240 V AC

BATTERY PACK 9780

NiMH, Charges while installed in the main unit

Storage media

*The CF card includes a PC card adapter. PC Card Precaution Use only PC Cards sold by HIOKI. Compatibility and performance are not guaranteed for PC cards made by other manufacturers. You may be unable to read from or save data to such cards.

 

PC CARD 512M 9728

• 512 MB capacity • PC card adapter bundled

PC CARD 9830

• 2 GB capacity • PC card adapter bundled

PC CARD 1G 9729

• 1 GB capacity • PC card adapter bundled

Is your radiant heating and cooling (HVAC) system producing the right amount of warmth? 

Measurement the heat generated by electronic components 

Drivers, Firmware, Software

Translate »